Your Position: Home > Brand > Endress Hauser

Proline Promag 10D Electromagnetic flowmeter

Update: 2016/5/31      View:
  • Brand:    Endress+Hauser
  • Type:    10D
  • Order Online
Introduction

The highly cost-effective flowmeter designed as compact wafer version

Promag D, the wafer device is designed for all applications where space is at a minimum. Combined with the Promag 10 transmitter for basic applications and direct integration, Promag 10D offers accurate measurement of liquids for a wide range of applications. It will be the preferred solution for customers in the water industry aiming for minimized cost of ownership. Promag 10D is available in a compact or remote version.


Benefits

  • Easy, fast centering of the sensor – innovative housing construction

  • Energy-saving flow measurement – no pressure loss due to cross-section constriction

  • Cost-effective – designed for easy applications and direct integration

  • Safe operation – display provides easy readable process information

  • Fully industry compliant – IEC/EN/NAMUR

  • Maintenance-free – no moving parts

Field of application

  • The measuring principle is virtually independent of pressure, density, temperature and viscosity

  • For basic water applications; optimized for limited space and plastic pipe installations.

Device properties

  • Short installation length and low weight

  • Integrated ground disks made of stainless steel

  • International drinking water approvals

  • 2‐line display with push buttons

  • Device in compact or remote version

  • HART

Measuring Principle Electromagnetic
Product headline The highly cost‐effective flowmeter, available as compact wafer. For basic water applications; optimized for limited space and plastic pipe installations.
Sensor features Easy, fast centering of the sensor – innovative housing construction. Energy‐saving flow measurement – no pressure loss due to cross‐section constriction. Maintenance‐free – no moving parts. Short installation length and low weight. Integrated ground disks made of stainless steel. 
Transmitter features Cost‐effective – designed for easy applications and direct integration. Safe operation – display provides easy readable process information. Fully industry compliant – IEC/EN/NAMUR. 2‐line display with push buttons. Device in compact or remote version.
Nominal diameter range DN 25 to 100 (1 to 4")
Wetted materials Liner: Polyamide 
Electrodes: 1.4435 (316L)
Measured variables Volume flow
Max. measured error Volume flow: ±0.5 % o.r. ± 2 mm/s (±0.5 % o.r. ± 0.08 in/s)
Measuring range 9 to 4700 dm3/min (2.5 to 1250 gal/min)
Max. process pressure PN 16, Class 150, 10K 
Medium temperature range 0 to +60 °C (+32 to +140 °F)
Ambient temperature range –20 to +60 °C (–4 to +140 °F)
Sensor housing material AlSi10Mg, coated 
Sensor connection housing: AlSi10Mg, coated
Transmitter housing material Powder‐coated die‐cast aluminum
Degree of protection Compact version: IP66/67, type 4X enclosure 
Sensor remote version: IP66/67, type 4X enclosure 
Transmitter remote version: IP67, type 4X enclosure
Display/Operation 2‐line display with push buttons 
Configuration via local display and operating tools possible
Outputs 4‐20 mA HART (active) 
Pulse/switch output (passive)
Inputs None
Digital communication HART 
Power supply DC 11 to 40 V 
AC 85 to 250 V (45 to 65 Hz) 
AC 20 to 28 V (45 to 65 Hz)
Hazardous area approvals FM
CSA 
Other approvals and certificates Calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025), NAMUR 
Drinking water approval: ACS, KTW/W270, NSF 61, WRAS BS 6920